Saving a configuration, Calibration – Cannon Instrument miniQV-X User Manual

Page 64

Advertising
background image

CANNON

®

miniQV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— October, 2008; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

58

Select which bulb to apply the Vent bulb drain time parameter.

Bulb overdraw

Use the Bulb Overdraw button/window to set the the amount of time in
seconds that the instrument will draw sample AFTER the sample menis-
cus has covered the target sensor in the viscometer tube. More viscous
materials may require a larger overdraw time to ensure that the sample
meniscus does not dip below the target sensor during temperature equili-
bration.

Proper adjustment of the overdraw can maximize efficiency of the test
cycle. Too much overdraw can contaminate vacuum lines and interfere
with pump operation. Too little overdraw can interfere with the initiation
of sample testing following the temperature equilibration period. Note
that lighter oils will be pulled higher in the bulb.

Saving a configuration

Whenever you have altered instrument, tray or sample configuration, you
should save the resulting settings as a new or existing configuration. To
save current settings, use the Save Instrument option:

Procedure

1. From the VISCPRO

®

primary menu options, click Configure.

2. Select the desired instrument and click Save Instrument & Tray

Settings. The Save Configuration window will appear.

3. Type a name for the new configuration in the Save As: box. Or click

an existing configuration from the configuration list box.

4. Click OK to save the configuration. Or click Cancel to exit the

Save Instrument & Tray Settings window without saving your
changes.

Calibration

The Calibration window provides controls for calibrating each bulb
associated with your instrument. You may need to log onto the system as
a Manager to access these security-protected calibration functions.

Use the bulb spin controls

to display current constants and valid

check standard data available for calibration of each bulb.

Calibration constants are determined individually for each bulb. Data for
automatic calculation of tube calibration constants is derived from check
standards run using that tube.

For information regarding the miniQV-X calibration procedure, see
Chapter 3, Calibrating the miniQV-X.

The Calibration window options are described briefly below:

Advertising