Cannon Instrument miniQV-X User Manual

Page 70

Advertising
background image

CANNON

®

miniQV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— October, 2008; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

64

with CANNON Instrument Company personnel if problems with sensor
training persist.

To set security levels, first log into the VISCPRO software as a Manager.
This will permit you to determine Manager or Technician accessibility
for the following protected menu-accessible functions:

Configuration:

Instrument Settings
Tray Settings: Tube & Bath
Tray Settings: Test
Tray Settings: Wash
Tray Settings: Advanced
Calibration

Service:

Adjust Vacuum
Temperature Calibration (saved as Instrument Settings)
Zero Temperature Offsets (saved as Instrument Settings)
Train Tube Sensors

To make security changes, click Service/Set Security Levels for the
desired instrument. The Service/Set Security Levels window will
open. Current security settings will be displayed. To change the security
level for a protected function, click on the Security Level cell opposite
the function name until the Manager/Technician drop-down box is
displayed. Then click the desired security level to select it.