The Calibration window

NOTE

You must click Update tube before changing bulbs if you wish to save
the new calibration constants for that bulb. If you close the
Calibration
window without updating the tube, the new calibration information will not
be saved.

Service menu

To access the service menu items, click Service from the VISCPRO

®

primary display and select the desired instrument.

The Monitor window permits the user to view current operational values
for the miniQV-X instrument. These settings may be helpful to CAN-
NON

®

technical personnel when troubleshooting difficulties with the

miniQV-X.

Select Wash viscometer tube to open the Wash viscometer tube
window.

Click Begin Wash to initiate the wash cycle. The Status: indicator will
display and update current actions.

Bypass .... option

To begin washing an evacuated sample cup, click the option box to make
certain that the Bypass initial cup evacuation option is selected.

CAUTION

Do not use this option with a filled cup or the cup will overflow.