Ventilation, Checking bath temperature, Dual-solvent washing – Cannon Instrument miniQV-X User Manual

Page 87: Solvent wash by computer

Advertising
background image

81

CANNON

®

miniQV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— October, 2008; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Ventilation

The miniQV-X requires installation of the supplied 3” ventilation hose to
the exhaust shroud on the rear panel of the unit. The shroud must be
connected to a customer-supplied exhaust fan/ventilation system. A 3”-4”
hose adapter is shipped from CANNON Instrument Company with each
miniQV-X unit. The air flow rate and design of the exhaust fan must be
suitable to handle the vapors from the customer-supplied solvent and
samples.

Checking bath temperature

After the miniQV-X has been turned on, you may monitor bath tempera-
ture via a reference thermometer in the bath or via the software tempera-
ture display (Sample Input window). Temperature should rise until the
bath has attained the desired test temperature. Consult the reference
thermometer to ensure that the miniQV-X is controlling at the desired
operating temperature. If the computer temperature display does not
agree with the reference thermometer, perform the temperature calibra-
tion procedure per manual instructions.

Dual-solvent washing

For testing heavy oils or hard-to-clean materials, CANNON

®

Instrument

Company recommends a dual-solvent washing system. The dual-solvent
washing process uses two different solvents. The first wash might
ordinarily use a heavy solvent to clean the sample residue out of the
viscometer tubes. The second wash might use a lighter solvent to clean
the remaining sample and solvent residue from the viscometer tube.
Consult Chapter 1 for information on dual-solvent connections.

Solvent wash by computer

Use the following procedure for computer-controlled solvent washing:

Procedure

1. If the desired tube is testing a sample, wait until the test is complete

OR click the Abort button from the Sample Input window and
select the tube/tray to abort from the Select Trays window. Then
click OK.

2. Proceed with the solvent wash by clicking the Wash Tubes button

from the Sample Input or Machine Status windows. Or you may
click on Service/Wash Viscometer Tube from the VISCPRO

®

primary menu options.

3. Click on Begin Wash to initiate a solvent wash using current wash

settings. During the wash cycle, the Begin Wash button will change
to Abort Wash. If it is necessary to interrupt the wash cycle, click
Abort Wash.

NOTE

The wash procedure will have to be successfully completed before
further sample testing will be allowed.

Advertising