miniQV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— October, 2008; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

62

To calibrate for temperature, see Chapter 3 for details. Temperature
calibration requires a reference thermometer with an accuracy meeting
the precision specification of ASTM D445.

To zero temperature offsets, click the Service/Clear Temperature
Offsets menu option for the desired instrument. The Zero Temperature
Offsets window will open. Click the check box to select each desired
action. Then click OK to clear the selected settings.

If the Clear Current Offset option is checked, VISCPRO will clear the
calibration point for the current bath temperature.