Cannon Instrument D155 Density Meter User Manual

Page 22

Advertising
background image

19

↓ [ENTER] Press [ENTER] key.

If the peristaltic pump is installed, the pump will run for the amount of seconds set in
the Menu:2 Sampling Time.

The above screen will only appear when using the optional peristaltic pump and
the Sampling Time (Menu:2 Sampling Time) is set to a value greater than zero.

7) When sampling is complete, the display shows:


Visually inspect the cell through the cell window to ensure that no bubbles are
present in the cell.

8) When calibration is normally completed, the display shows:


When “NG” (No Good) appears instead of “OK”, press [STOP] and then repeat
the calibration starting with step 3 above.

9) If the optional peristaltic pump is installed, press [PUMP] key to drain out water.

Otherwise use an empty syringe to push out the water with air.

10)

Blot the sample drops on the sampling tube with tissue paper.

Note:

If calibration fails, previously calibrated values will be printed out.

M e n u : 0

C a l i b r a t e C e l l

S A M P L I N G : 1 0 s

< < P l e a s e W a i t > >

M e n u : 0

C a l i b r a t e C e l l

C A L I B R A T I N G

< < P l e a s e W a i t > >

M e n u : 0

C a l i b r a t e C e l l

O K

P r e s s S T O P o r M E N U

CANNON

®

D155 Density Meter Instruction & Operation Manual

Version 1.1b – August 2014; CANNON® Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Advertising