4-2. Menu 2: Set Sampling Time

Select sampling time adequate for the sample to be injected. The default is ‘10’ (seconds).

1) Press [MENU] then press [] or [] to select Menu 2 as shown below

2) Press [ENTER]

3) Press [] or [] key to select the digit to change
4) Press [] or [] to increase or decrease the number. The D155 allows any value from 00 (pump

off or disabled) to 99 seconds. The default setting for instruments with the pump installed is 10.

5) Press [ENTER] to accept the value entered


4-3. Menu 3: Select Oil Table

The API temperature compensation table for oil can be selected as follows. It depends on the samples
you wish to measure.

1) Press [MENU] then press [] or [] to select Menu 3 as shown below

2) Press [ENTER]

3) Press [] or [] to select the appropriate API temperature compensation table. The available

tables are:

23A

Crude Oils 60/60°F

23B

Refined Products 60/60°F

23D Lube Oils 60/60°F
53A

Crude Oils 15°C

53B

Refined Products 15°C

M e n u : 2

S e t S a m p l i n g T i m e

P r e s s E N T E R o r U p / D n

M e n u : 2

S a m p l i n g T i m e : 1 0 s

A r r o w K e y s t o C h a n g e

E N T E R t o A c c e p t

M e n u : 3

S e l e c t O i l T a b l e

P r e s s E N T E R o r U p / D n

M e n u : 3

O i l T a b l e : 5 3 A

A r r o w K e y s t o C h a n g e

E N T E R t o A c c e p t