6-3-2. “Ready to Test” message does not appear and/or cell temperature is unstable.

Follow the following steps if “Ready to Test” message does not appear after more than 1 hour or the cell
temperature appears unstable after having reached the preset measurement temperature.

START

Check the air-intake at the

bottom to see if clogged

Check ambient

temperature (5°C~+35°C)

Operate the unit within the range

of the specified temperature

No

Yes

Check the air-intake at the bottom

to see if clogged

No

Yes

END

Repairing by manufacturer

Is everything O.K.?

No

Yes

6-3-3. The unit does not work even though power is on.

If the “Ready to Test” message does not appear after the D155 has been powered on for more than 30
minutes, check on the line voltage. If correct power voltage is supplied, Contact CANNON or your local
dealer.

6-3-4. LCD does not function

If the LCD display is found to be defective (no message appears on screen) or the brightness of display
cannot be adjusted, Contact CANNON or your local dealer.