35

5-2-3. Replacement of optional peristaltic pump tubing

If the pump tube is deteriorated, the tubing may fail causing sample liquid to leak. This may eventually
lead to pump failure. Replace pump tubing as follows:


1) Disconnect the tubes connected to the pump tube cassette.
2) Turn the cassette counterclockwise to

remove it.

3) Remove the tube joint from the cassette by

pulling one of the ends, and then pull out the
tube.

4) Apply grease (such as petroleum jelly) over

the tube and the axle of pump cassette as
illustrated.