2. principle of measurement, 2-1. oscillating u-tube cell – Cannon Instrument D155 Density Meter User Manual

Page 48

Advertising
background image

45

7-2. Principle of measurement

7-2-1. Oscillating U-tube Cell

When a measurement cell filled with liquid or gas is oscillated spontaneously, resultant oscillating cycle
varies with the density of the sample in the measurement cell. The oscillating cycle T is expressed
follows.

T

dVc Mc

K

=

+

where,
d = Density of sample in measurement cell
Vc = Volume of sample in measurement cell, or measurement cell internal volume
Mc = Mass of measurement cell
K = Constant

d

K

Vc

T

Mc

Vc

=

4

2

2

π

Supposing that the oscillating cycles Ta and Tw are obtained through measurement of reference
materials of known density (density da and dw), the facto value F is calculated as follows.

F

K

Vc

d

d

T

T

a

w

a

w

=

=


4

2

2

2

π

Accordingly, the density “d” of unknown sample can be calculated according to the following formula by
measuring its oscillating cycle T.

d d

F T

T

a

a

=

(

)

2

2


CANNON

®

D155 Density Meter Instruction & Operation Manual

Version 1.1b – August 2014; CANNON® Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Advertising