Cannon Instrument D155 Density Meter User Manual

Page 24

Advertising
background image

21

5) After measurements are completed, the results will appear on the screen

display.


If a printer is connected, the results will automatically print.

There are four results display windows. To change between these
windows, press the [] [] keys. Below are the three additional results
displays.

6) When continuing measurement on the same sample, return to Step 1).


When the intended number of measurements is completed, remove the
sample from the cell. If the optional peristaltic pump is installed, press
[PUMP] key to drain out the sample. Otherwise use an empty syringe to
push out the sample with air.

7) Clean the measuring cell according to Section “3-4. Clean measuring cell

after measurements”.

Note:

When you have changed the measurement temperature, make sure to

perform calibration before measuring at the new temperature. The D155

will not allow a measurement to be made at a temperature which has not

been calibrated.

1 0 / 3 0 / 1 2 0 9 : 5 0

D e n s i t y 0 . 9 9 8 5 g / c m ³

T e m p 2 0 . 0 0 ° C T e s t 0 1

P r e s s U p / D n U n i t s

1 0 / 3 0 / 1 2 0 9 : 5 0

A P I D e g r e e 5 D : 2 4 . 1

T e m p 2 0 . 0 0 ° C T e s t 0 1

P r e s s U p / D n U n i t s

1 0 / 3 0 / 1 2 0 9 : 5 0

1 5 ° C D e n 5 3 D : 0 . 9 9 9 5

T e m p 2 0 . 0 0 ° C T e s t 0 1

P r e s s U p / D n U n i t s

1 0 / 3 0 / 1 2 0 9 : 5 0

S G ( t / t ) : 0 . 9 9 9 0

T e m p 2 0 . 0 0 ° C T e s t 0 1

P r e s s U p / D n U n i t s

CANNON

®

D155 Density Meter Instruction & Operation Manual

Version 1.1b – August 2014; CANNON® Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA