6-3-7. There is poor repeatability on measured values

Contact our local dealer to

check the cell.

START

Is the sample homogeneous?

Is there

any repeatability for the

measurement with

pure water?

Is the work bench vibrating?

Any sample residue

in joint portions (tube nozzle A,

joint B or the like)?

Are air bubbles found

in the sample?

Is the measuring cell dirty?

Is the temperature of

measuring cell stable?

Repeat sampling. (When air bubbles

appear again, refer to Section 6-3-6.)

Stir to make it homogeneous. (No

repeatability when the sample is

unevenly mixed.)

Overhaul the joints and clean

them.

Rinse the cell thoroughly

.

Measure after temperature

becomes stabilized.

Yes

No

Contact our local dealer.

Is everything O.K.?

END

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

Measure the sample on a

stable bench.

No

CANNON

®

D155 Density Meter Instruction & Operation Manual