Row window in plot manager – ChemoMetec NC-250 User Manual

Page 33

Advertising
background image

991-0252 Rev. 1.1

February 2013

33


Right-click in the table offers an option to copy table contents to clipboard which can be directly pasted
into a spreadsheet program.

Row Window in Plot Manager

The Plot Manager Row Window is used as row header for the post processing of a specific image file. It is
always placed in the left side of the Plot Manager window.

Row number: The row number is shown in the upper left corner.

Data source indicator: The data source of the cell population in this row is shown with a small colored
square to the left of the ‘save’ button:

1. Grey: Indicates that the cell population has been loaded from data embedded in the image file, i.e.

data which was obtained when the image was originally acquired and analyzed by running a
protocol.

2. Green: Indicates that the data has been loaded from the main window, and that the main window

still holds the image file in memory. This is referred to as Main data (synchronized). This population
is loaded instead of embedded data if the main window holds cell population data for that
particular image file at the time when the row is added to the Plot Manager.

3. Orange: Same as for green, but the main window no longer holds the same image file in memory.

This is referred to as Main data (not synchronized).

The ‘save’ button saves the post processing results in a *.cmpp file which is automatically given the
same name and same location as the *.cm image file.
Note: Any already existing post processing result for this CM-file will be overwritten. In 21CFR Part
11 mode only one saving of post processing results is allowed.