12

Material
Substitution:

When certified test weights are not available
you can use an accurately weighed material
instead.

The material must be weighed on a second-
ary, calibrated scale and delivered to the site
of the primary floor scale for calibration.

The secondary calibrated scale should be of
the same accuracy or greater and have a
capacity approximately equal to the primary
floor scale.

SCALE VERIFICATION

Step 1.

Get a test weight that is equal to 10% of
the full scale capacity.

Step 2.

Place the test weight at Position #1 on the
platform. (See Fig. 10)

FIG. 10 SCALE VERIFICATION/POSITION #3

Step 3.

Write down the weight for position #1.

Step 4.

Place the weight at position #2.

Step 5.

Write down the weight for position #2.

Step 6.

Place the weight at position #3.

Step 7.

Write down the weight for position #3.

Step 8.

Place the weight at position #4.

Step 9.

Write down the weight for position #4.

Step 10.

Check all the readings. If all the readings
are within an acceptable tolerance, no re-
calibration is required.

Position #3

Position #4

Position #2

Position #1