Hardy Floor Scales

37

Check for loose connections or broken wires.

Check the wiring color code to be sure that
the wires are routed correctly. If they are not
wired correctly change the wiring until it is
correct.

Check the Weighing Instrument Manual for
trouble shooting instructions.

Problem:

Scale indication is not linear.

Cause #1:

Packing material or debris wedged or built up under-
neath the platform

Remedy #1:

Carefully lift the platform high enough to
safely remove the built up debris.

Check for loose connections or broken wires.

Cause #2:

Weighing instrument or floor scale is not calibrated.

Remedy #2:

Re-calibrate either the weight instrument or
the floor scale or both.

Check for corrosion on the electrical connec-
tions.

Problem:

The scale reading drifts or is erratic.

Cause #1:

Corrosion or moisture in the electrical connections.

Remedy #1:

Check to see that the summing junction box
is tightly fastened.

Check to see that the gaskets and seals are
not cracked or damaged.

Remove the wires and clean any corrosion
from the connectors and exposed wire.

For high humidity areas, place a desiccant
packet in the summing junction card enclo-
sure.

Cause#2:

High voltage wires close to the load sensor cable.

Remedy #2:

Move the high voltage wires away from the load sen-
sor cable.

Cause #3:

Instrument or floor scale improperly grounded (i.e.
one or both not connected to earth ground).