Step 3.

Use a small slotted head screw driver and
disconnect the Load Sensor cable. It is a
good idea to mark the wires if the wire
markers have been removed. Color code
wires are as follows:

We recommend the use of C2 cable from the
weighing instrument to the junction box. Note the
color codes for C2 load points (left to right facing the
rear panel). Match the load sensor’s color coding to
ease trouble shooting.

Load Sensor #

Connector

Load Sensor #1

TB1

Load Sensor #2

TB2

Load Sensor #3

TB3

Load Sensor #4

TB4

To Instrument

TB5

Table 3: Summing Card Load Sensor Connectors

Load cell connector

Rear panel 9 pin

C2 cable TB9 7 pin

to Junction Box

Load cell color codes

3 pin connector

Shield Ground Wire

C2- Violet

C2+ Grey

EXC- Black

SEN- Brown

SIG- White

SIG+ Green

SEN + BLUE

EXC+ RED

EXC- Black
SEN- Brown
EXC+ RED
SEN+ BLUE
Shield Ground Wire
C2+ Grey
C2- Violet
SIG+ Green
SIG- White

EXC+ RED

EXC- Black

Shield Ground Wire

4 pin connector

C2+ Grey

C2- Violet

SIG+ Green

SIG- White