Chapter 5 - Operating Procedures

5-7

Step 6.

Set the ROCTimeBase - Click on the up
or down arrow to reach the value you want
or double click in the field and enter the
value. The Rate of Change Time Base is
the time period you want for weight sam-
pling to determine the Rate of Change,
either loss in weight or gain in weight. The
range is 1 - 1800 seconds.

Step 7.

Set the Zero Tolerance - Double click in
the Zero Tolerance field. The default value
is 10 units of measure. Enter the new Zero
Tolerance value. The range is 1 to 32766.
Zero Tolerance sets the range of weights
so that the Zero Command works, as an
offset of the calibrated Zero. Zero Toler-
ance is the number of graduations from
zero that will be accepted as zero by the
module. To enable Zero Tolerance click in
the check box next to “Zero Enabled” so
that a check mark is visible. The operator
can reset this value at any time.

Step 8.

Set the Auto Zero Tolerance - The
default value is 10 units of measure. The
range is 1 to 32766.

Double click in the Auto Zero Tol-
erance field.

Enter a positive value that does not
exceed the maximum zero tolerance
value or is not a negative value.

Click in the check box next to ”Zero
Track Enabled” to activate Auto
Zero Tolerance tracking.

The Auto Zero Tolerance setting
automatically sets the Zero Toler-
ance so that the Zero Command
works, as an offset of the calibrated
Zero.

Step 9.

Set the Motion Tolerance Value - Click
in the Motion Tolerance field and enter the
motion value. The motion tolerance is the
tolerance value used to determine if the
scale is in motion. If the motion is outside