Chapter 5 - Operating Procedures

5-9

3.

The deadband should be numerically larger than
the preact to prevent relay chatter.

Step 2.

Double click in the Value field.

Step 3.

Enter the desired set point value.

Step 4.

Double click in the Preact field.

Step 5.

Enter the desired Preact value. Preact is
the number of units above or below the set
point value of which the relay will trip.
Use as an “in flight” compensation value.

Step 6.

Double click in the Deadband field.

Step 7.

Enter the desired Deadband value. Dead-
band is a value used to prevent relay chat-
ter once the setpoint is reached. You can
use positive or negative values. The nega-
tive value can be used as a low trip limit,
the positive value can be used as a high
trip limit.

Step 8.

Click in the radio button next to the Set-
point Mode you want.

Step 9.

Click in the radio button next to the Value
units you want to read the weight. Gross,
Net, or ROC.

Step 10.

Click on the Save button to save the set-
tings.

Reading C2 Load Cell
Serial Numbers

Step 1.

Click on the C2 Tab. The C2 page
appears. (See Fig. 5-7)

Step 2.

Go to the Hardy Internet Site (http://
www.hardyinst.com) where you can look
up the parameters of your load cell(s).

Advertising