Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 26

Advertising
background image

EN-24

5000 Series Indicators

3.5.3 Filter

Set the amount of signal filtering.
LOW = less stability, faster stabilization time (<1 sec.)
MEd = normal stability, stabilization time (<2 sec.)
HI = greater stability, slower stabilization time (<3 sec.)

3.5.4 Auto-Zero Tracking

Set the automatic zero tracking functionality.
OFF

= disabled.

0.5 d = the display will maintain zero until a change of 0.5 divisions per second has been
exceeded.
1 d = the display will maintain zero until a change of 1 division per second has been
exceeded.
3 d = the display will maintain zero until a change of 3 divisions per second has been
exceeded.

NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d, 1d and 3d. The setting is
locked when the hardware lock switch is set to the ON position.

3.5.1 Reset

Set the Readout menu to factory default settings.
NO

= not reset

YES = reset
If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level, Auto Zero Tracking, Auto Off
and Gross settings are not reset.

3.5.2 Stable Range

Set the amount the reading can vary before the stability symbol turns off.
0.5d = 0.5 scale division
1d = 1 scale division
2d = 2 scale divisions
3d = 3 scale divisions
5d = 5 scale divisions

NOTE: When LFT is set to ON, the setting is forced to 1 d. The setting is locked when the hardware lock
switch is set to the ON position.

Advertising