Holtgreven Ohaus Defender Series Bench Scales User Manual

Page 41

Advertising
background image

EN-39

5000 Series Indicators

Contact
Set the timing of the relay contacts.

SIM = relays open or close at the same time.
b-b-M = relay opens before the next relay closes (break before make).
M-b-b = relay closes before the next relay opens (make before break).

NOTE: A 100 ms delay or over-lap is used for the break-before-make and make-before-break timing.

Stable
Set how the relay outputs react during instability.
OFF

= relay changes are immediate.

ON = delays relay changes until weight reading is stable.

3.11.5 End I-O

Advance to the next menu.

3.12 Menu Lock Menu

Reset No, Yes
Lock Calibration Menu Off, On
Lock Setup Menu

Off, On

Lock Readout Menu Off, On
Lock Mode Menu

Off, On

Lock Unit Menu Off, On
Lock Print1 Menu Off, On
Lock Print2 Menu Off, On
Lock Com1 Menu Off, On
Lock Com2 Menu Off, On
Lock GMP Menu Off, On
Lock I-O Menu Off, On
End Lock Menu

Use this menu to prevent unauthorized changes to menu settings. When the security
switch is set to ON, the locked menus can be viewed but not changed. Default settings
are bold.

3.12.1 Reset

Set the menu Lock menu to factory defaults.
NO

= not reset.

YES = reset.

NOTE: Settings for LFT controlled menu items are not reset.

3.12.2 Lock Calibration

Set the status.
OFF = Calibration menu is not locked.
ON

= Calibration menu settings is locked.

Advertising