IDS440 Users Manual version 4.2

Set Operation Parameters

25

Scale Parameters

The scale parameters configure IDS440 for the weigh platform or load cell that it is

connected to. The preceding table titled “SCALE PARAMETERS” lists the scale

parameters and their factory settings. Use the “Field Setting” column to record any

changes made to the factory default values.
Many of the following parameters are entered in terms of “graduations”. The full-scale

graduations can be found by dividing the full-scale weight by the count-by weight. For

example, if the scale is configured for 500.00kg x 0.05kg, the full-scale graduations is

500.00/0.05 = 10,000 graduations to full scale.
The numbers in parenthesis are the default factory settings.

Parameter 1. Load Cell mv/V (3 mv/v)

The IDS440 can be used with load cell outputs of 1 mv/V, 2 mv/V, or 3 mv/V. Enter a 1,

2, or 3 for the load cell type used in your application.

Setting

A/D gain

0

200

1 mV/v

100

2 mV/v

100

3 mV/v

50

4

25

Parameter 2. Digital Filter (9)

The digital filter removes scale vibrations and other noise signals from the weight

display. The filter range is between 0 and 18. A setting of 0 provides no filtering; a

setting of 18 provides maximum filtering. The higher the filter number, the longer the

settle time of the weight display. For example:

Filter Setting

Settle Time

3

0.4 sec

6

0.8 sec

9

1.75 sec

12

3.5 sec

15

7.0 sec

Use the smallest filter setting that provides satisfactory display stability.

Parameter 3. Display Update Rate (1)

Enter the time between display updates (in tenths of a second). An update rate of 1

causes a 0.1 second delay between display updates, a rate of 200 causes a 20 second

delay between updates. NOTE: motion detection is calculated using the weight values of

the display rate. If the update rate is 50, the motion calculation is based on weight

readings sampled at 0.5sec intervals.