Set Operation Parameters

29

Parameter 21. Deadload Factor

The deadload factor is automatically calculated by the calibration function. This

parameter is not accessible from the keyboard.

Parameter 22. Weight Conversion Factor

The weight conversion factor is automatically calculated by the calibration function.

This parameter is not accessible from the keyboard.

Serial I/O Ports

The IDS440 is equipped with two bi-directional serial communication ports that can

communicate to an optional printer, a computer, bar code reader, a score board or other

serial communication devices. The baud rate and format can be set independently for

each port. Serial port 1 is normally used to output weight data and ID reports to a

printer. Serial port 2 is normally used for continuous output of weight data. Continuous

transmission can be enabled or disabled and the output format can be customized using

Parameters 27 and 28.
Communication between serial devices requires that both devices have the same baud

rate (transmission speed), data bit length, and parity. The mode parameter selects the

data bit length and parity the baud parameter selects the baud rate.

Parameter 23 Serial Port 1 Mode (4 [8 data bits, no parity])

Enter the mode number from the table below to select the mode parameter in the mode

description column.

Mode Number

Mode Description

1

7 data bits, no parity

2

7 data bits, even parity

3

7 data bits, odd parity

4

8 data bits, no parity

5

8 data bits, even parity

6

8 data bits, odd parity