IDS440 Users Manual version 4.2

Parameter Functions

38

3. Press the ‘1’ key to adjust by 1, press the ‘2’ key to adjust by 10, the ‘3’ key to adjust

by 100, or the 4 key to adjust by 1000. Use the UNITS key (

↓) and the PRINT key

(

↑) to increase or decrease the displayed weight until the displayed weight is the same

as the weight on the scale. The center of digit indicator (-) should be displayed for

maximum accuracy.

4. Press the CLEAR key to exit Adjust Gain procedure.

Adjust Gain – 5 Point Linearization

If 5 point Linearization is enabled, the gain adjust procedure begins with the display

alternating between the calibration number (example: Cal 1) and the calibration weight.
• Press the ENTER key to adjust the gain at the displayed calibration point.
• Press the UNITS key (↓) to adjust the next calibration point
• Press the PRINT key (↑) adjust the previous calibration point.
• Press the ZERO key to clear a calibration point. (Must re-calibrate to reuse).

Function 64 Configure Passwords

Use function 64 to enter password data. The IDS440 provides two security passwords

that can be activated and entered to protect configuration data. Password 1 is used to

protect I/O port configuration, setpoint, and event monitor data. Password 2 is used to

protect calibration and scale configuration data. Security level 3 provides extra

protection for password level 2 data. Security level 3 requires the entry of password 2

and the Function Lock inside the IDS440 to be in the unlocked position. Password 3 will

inhibit viewing, changing, and uploading of Scale Basic functions to a PC when

password is active but will allow downloading a new Scale Basic program from the PC to

the indicator. All passwords will be disabled once the new program has been downloaded

into the indicator and should be reactivated.
The Operation Parameter Reference Table lists the configuration parameters and the

password/security level required to access them. The IDS440 prompts “PASS 1” when a

parameter is selected that requires password level 1. If the entered password is incorrect,

the IDS440 displays “Error”. Press the ENTER key to try again. Press the CLEAR to

exit from password entry.
IMPORTANT NOTE: Write down the passwords and put them in a safe place. If you

loose and forget a password, there is no way to access the protected functions.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.

2. Enter CFG 64. The IDS440 prompts “P1 xxxx” where xxxx is the current password

number. Enter a new password number or press the ENTER key to use the displayed

number. Enter a 0 (zero) for the password number to disable the password.

3. The IDS440 prompts “P2 xxxx”. Enter password 2 data.

4. The IDS440 prompts “P3 xxxx”. Enter password 3 data.