AD-4407 Weighing Indicator

Note that the

ON/OFF

key does not function alone. Press the

ESC

key while

holding the

ON/OFF

key to end the calibration mode, if mis-operation. After

displaying Can5el, press the

ON/OFF

key to stop calibration mode and to

turn the indicator off.

NOTE: When displaying Caloff , press the +/- key while pressing the

ON/OFF

key,

instead of

CAL

key.

Caution

The maximum display is less than or equal to 40000 divisions. This

number is calculated from the maximum capacity divided by the

minimum division.

Check the accuracy of weighing instrument periodically.

Recommended mass, use a mass heavier than 2/3 maximum capacity.

Calibrate the scale, if it is moved to other location or the environment

has changed.

It is not necessary to input the gravity acceleration correction, when

calibrating the scale with a calibration mass at the place where the

scale is used.

Enter the stable weighing data while the STABLE mark is displayed. If

unstable data is used, it may cause a weighing error. Arrange the

condition using the F00 filter function.

The span calibration needs the zero calibration data. We recommend

that you perform the span calibration immediately after the zero

calibration.

If you use the dual range function of the multi-interval scale, perform

the "Range Function", "Zero Calibration" and "Span Calibration".

5.2.

5.2.

5.2.

5.2. Calibration

Procedure

Calibration Procedure

Calibration Procedure

Calibration Procedure

5.2.1.

Configuring a Weighing Instrument

This section explains how to specify capacity, resolution, decimal point position and

display format, weighing range and unit. Perform this procedure when installing the

indicator.

When Cal5et is displayed, enter this mode by pressing of the ENTER key.

Specify the range and unit.