7. tare tare tare tare – A&D Weighing AD-4407 User Manual

Page 33

Advertising
background image

AD-4407 Weighing Indicator

Page

31

7.

7.

7.

7. Tare

Tare

Tare

Tare

The function is used to display a net value with the container weight subtracted from

the total weight, if you place an object into a container to weigh it.

Using a serial interface such as the RS-232C, you can do this from the external

equipment.

Caution

When turning the display off with

Cf02 1

Cf02 1

Cf02 1

Cf02 1, 2222 or 3333, the tare data is cleared.

When turning the power off, the tare data is cleared.

Weighing Tare

Weighing Tare

Weighing Tare

Weighing Tare

Operation

Place the tare on the weighing unit. Press the

TARE

key to store the tare

weight after the STABLE mark turns on. The display changes to net.

Caution

When displaying a negative gross value, tare can not be used(normally).

To enable tare at zero or a negative gross value, select a

Cf 04

Cf 04

Cf 04

Cf 04 parameter.

Digital Input

Digital Input

Digital Input

Digital Input

Caution

The input value is rounded off to the unit of division (weighing interval).

In the case of

Cf02 1

Cf02 1

Cf02 1

Cf02 1, 2222 or 3333(power-on zero), the displayed value will be

zeroed when turning the display on.

When using a multi-interval scale, the usable input range is the first range.

In the case of

Cf09 1

Cf09 1

Cf09 1

Cf09 1(To inhibit preset tare), preset tare can not be used.

Preset tare is stored as one of the code memory data. Refer “9. Code

Memory”.

Preset tare value can be set via the serial interface.

Step 1

Press the PT key to display the stored tare value.

When tare is cleared or is not used, the value is zero.

Step 2

Enter a new tare value using the

0

-

9

keys.

ENTER

key

The key to store a new tare value.

The net is displayed.

ESC

key

The key to return to weight display without changing the stored

value.

Clearing Tare

Clearing Tare

Clearing Tare

Clearing Tare

When pressing the

TARE

key while gross is zero, tare is cleared and gross is

displayed.

When zeroing with the

ZERO

key, tare is cleared.

7.

Tare

Advertising