Installation and precautions, The load cell connections – A&D Weighing AD-4407 User Manual

Page 7

Advertising
background image

AD-4407 Weighing Indicator

Page

5

3.

3.

3.

3. Installation and Precautions

Installation and Precautions

Installation and Precautions

Installation and Precautions

3.1.1.

Installation and Precautions

The weighing indicator is a precision electronic instrument. Handle it carefully.

The operating temperature is -10°C to +40°C (14°F to 104°F).

Do not install the scale in direct sunlight.

Mis-operation or other problems may be caused by an unstable power source including

momentary power failure or instantaneous noise. Use a stable power source.

Do not connect the power cord before the installation has been completed.

Verify that the local voltage and receptacle type are correct for your scale.

Use shielded cable for all connections. Connect the cable shields to the shield

terminal or case as an earth terminal.

Earth ground the indicator. Do not join the earth ground line with other electrical

power equipment. (Example: There is an earth ground terminal at the power cord receptacle.)

Do not install the indicator in a place where it is apt to be charged with static

electricity, or where the relative humidity is lower than 45%RH. Plastic and insulating

materials are apt to be charged with static electricity.

3.1.2.

The Load Cell Connections

Connect the load cell cable to the terminal as shown below.

It is possible to connect a 4 wire cable provided that pins 1-2 and pins 3-4 are

shorted, if the distance between the indicator and a load cell is shorter than 5m.

The output voltage of a load cell is a very sensitive signal. Space the load cell cable

away from any noise source.

It is possible to connect eight 350ohm load cells.

The load cell drive is 5VDC 5% between EXC+ and EXC-, the maximum current

120mA.

3. Installation and Precaution

Loadcell

Cable

Positive Signal In

Negative Signal In

Negative Sense

Negative Excitation

Positive Excitation

Positive Sense

Shield

Standard connection

1

2

3

4

5

6

7

Cable

Shield

Available connection for

a cable shorter than 5m.

1

2

3

4

5

6

7