AD-4407 Weighing Indicator

Specifying the Weighing Range of the First Range

Specifying the Weighing Range of the First Range

Specifying the Weighing Range of the First Range

Specifying the Weighing Range of the First Range

Step 3 After displaying Cap for 2 seconds, single range or the weighing capacity will be

displayed. When dual range is used, Cap1 is displayed for 2 seconds. Triangle

1 will be displayed. Specify data with the

0

-

9

keys and press the

ENTER

key to

store it and proceed to the next step. When pressing the

ESC

key, regardless of

what is displayed, the indicator will proceed to the next step without changing it.

The next step is Zero Calibration in single range or the second range resolution in

dual range.

Specifying the Second Range Resolution

Specifying the Second Range Resolution

Specifying the Second Range Resolution

Specifying the Second Range Resolution

Step 4 After displaying range2 for 2 seconds, the resolution with decimal point and

triangle

2 will be displayed. Specify the second range resolution in the same

way as the first range. The decimal point cannot be moved. Specify the second

range resolution greater than the first range. Press the

ENTER

key to store them

and proceed to the next step. When pressing the

ESC

key, regardless of what is

displayed, the indicator will proceed to the next step without changing them.

Specifying the Second Range Capacity

Specifying the Second Range Capacity

Specifying the Second Range Capacity

Specifying the Second Range Capacity

Step 5 After displaying Cap2 for 2 seconds, the capacity with unit and decimal point is

displayed. Specify the capacity in the same way as the first range. The capactity

should be greater than the first range. Press the

ENTER

key to store it and proceed to

Zero Calibration. When pressing the

ESC

key, regardless of what is displayed, the

indicator will proceed to the next step without changing the parameter.

5.2.2.

To Get Stabilized Data

Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.

Maintain a constant temperature, stable power and stable input voltage from the

load cell.

Avoid direct sunshine or the near the outlet of an air conditioner.

Do not install the scale (indicator) where there is a strong magnetic field.

Step 7 Turn the display on and leave it for several minutes.

5.2. Calibration Procedure