A&D Weighing AD-4407 User Manual

Page 39

Advertising
background image

AD-4407 Weighing Indicator

Page

37

10.1.1.

Condition Formula for Comparison

Comparison is performed based on the following formula.

Upper/lower comparison

Judge Condition

Formula

Display Output

HI

Upper limit (Hi limit) value < Displayed value

HI

HI

OK

Lower limit value

Displayed value Upper limit value

OK

OK

LO

Displayed value < Lower limit (Lo limit) value

LO

LO

5-stage comparison

Judge Condition

Formula Display

Output

HiHi

HiHi limit value < Displayed value

HI, 2

HI

HI

Upper limit (Hi limit) value < Displayed value

HI

HI, OK

OK

Lower limit value

Displayed value Upper limit value

OK

OK

LO

Displayed value < Lower limit (Lo limit) value

LO

LO, OK

LoLo

Displayed value < LoLo limit value

LO, 3

LO

The decimal point is not considered. Example: If the upper limit value is 10.0, enter 100.

These parameters are stored in non-volatile memory, and are not lost even if the

AC power is disconnected or cut off (except code 0 memory).

When the displayed value becomes an overload (positive over), HI (over) is output.

When the displayed value becomes an under load (negative over), LO (under) is output.

This function compares the Hi / HiHi limit value first.

This function does not check the relationship between the upper and lower limit values.

10.1.2.

Setting the Upper/Lower Limit Values

Step 1 By pressing the

SETPOINT

key, the selected code memory number is displayed

and the first comparison class term (ex.Hi, TG etc.) is blinking.


Step 2 Select the comparison class using the following keys:

SETPOINT

key ........The key to select a comparison class.

ESC

key ..................The key to return to the previous stage (weight display or

code memory number select).

ENTER

key..............The key to proceed to step 3.

Step 3 Set the setpoint values using the following keys.

0

-

9

key..................The key to enter a value.

ESC

key ..................The key to proceed to the next comparison class (step 2)

without changing the set value.

ENTER

key..............The key to store the value and proceed to the next

comparison class (step 2).

10.1. Weight Check Mode

Advertising