Calibrate

When black trap calibration is completed, the
instrument must be placed on the white standard.

Position the instrument as thoroughly on the white
standard
as it is described for the black trap and
press the operate button.

The reading process is indicated on the display.

When calibration is complete, an affirmation is
shown in the display. Confirm by pressing the
Operate button or the wheel to proceed to the Cyan
check.

Choose "OK" with the scroll wheel.

When the cyan check is completed, an affirmation
is shown in the display. Confirm by pressing the
operate button or the wheel to return to the main
menu.

If an error occurs during calibration, try to
read the plaque again. If the error message

still occurs, refer to section "Errors and warning
messages“.