14. Technical Data

General technical data:

Temperature range

10 °C to 42 °C (50 °F to 110 °F)
for operation
0 °C to +60 °C (32 °F to 140 °F )
for storage

Rel. humidity

Up to 85% non-condensing/35°C (95 °F)

Measurement unit:

Color
Measuring Geometry

Illumination 45°,
-15°,15°, 25°, 45°, 75°, 110° aspecular viewing

Measuring Area

Ø23mm (0.91 in.) or 12mm (0.47 in.)

Spectral Range

400 - 700 nm, 10 nm resolution

Measurement Range

0 to 400 % reflectance

Measuring time
BYK-mac

< 6 seconds

BYK-mac COLOR

< 4 seconds

Repeatability

0.02 dE*, 1
10 consecutive readings on white tile

Reproducibility

0.2 dE*, 1
average on 12 BCRA II tiles

Color scales

dE*, dE CMC, dE 94,dE DIN 6175

Illuminants

A; C; D50; D65; F2; F7; F11; F12

Observer

2°; 10°

Effect

only with BYK-mac

Measurement Geometry 15°, 45°, 75° and diffused illumination

perpendicular viewing

Effect parameters

dS, dS_a, dS_i, dG

Repeatability

S_a / S_i: 5% or > 0.50 / G = ± 0.05, 1

Reproducibility

S_a / S_i: 10% or > 1.00 / G = ±0.15, 1

Memory

1000 Standards/Samples

Power supply

Rechargeable battery pack or 4 mignon AA batteries
(alkaline or rechargeable)

Dimensions (LxWxH)

218 x 81 x 147mm (8.6 x 3.2 x 5.8 in.)

Weight

approx. 1300g/ 2.86 lbs

Interface

USB-B