ALTANA BYK-mac COLOR Manual User Manual

Page 24

Advertising
background image

23

6.3 Effect Check

Choose "Absolute" from the "Measure" menu. The
basic configuration is the average of three
measurements.

Open the cover of the effect checking standard and
position the instrument on the standard. The
instrument snaps into the indentation of the
standard and can only be positioned in one
direction.

Perform three measurements.

Compare the average of the measurements with the
range indicated on the lid.

If the values measured on the effect standard are
within the printed tolerance range, the requirements
are met.

If the mean value is not within the desired tolerance
range, try carefully cleaning the effect standard. If
this produces no improvement, please contact our
Customer Service department.