Operation

BYK LC 4

26

4.5

Setting the temperature compensation

The calculation of the temperature compensation is based on the preset
reference temperature, Tref20 or Tref25 (see section 4.6 S

ETTING

THE

REFER

-

ENCE

TEMPERATURE

).

You can select one of the following temperature compensation methods:

! Nonlinear temperature compensation (nLF)
! Linear temperature compensation (Lin) with selectable coefficients of

0.001 ... 3.000 %

/

K

! No temperature compensation (- - - -)

Note
Select the following temperature compensations given in the table according
to the respective test sample:

Application tips

Measuring solution

Temperature compensation
(TC)

Display
indicator

Natural water (ground
water, surface water
and drinking water)

nLF
according to ISO 7888, EN
27888

(*)

(*) Corresponds to DEV C8

Ultrapure water

nLF
according to ISO 7888, EN
27888

(*)

Other aqueous solu-
tions

Set linear
temperature coefficient 0.001
... 3.000 %/K

Salinity
(seawater)

Automatic nLF according to
IOT

Sal,