4 transport, 5 disposal, Transport – ALTANA LC 4 User Manual
Page 45: Disposal
BYK LC 4
Maintenance, cleaning, disposal
45
Cleaning the
sensor module
We recommend to clean the sensor module thoroughly, especially before
measuring low conductivity values. Before measuring, thoroughly rinse the
sensor with deionized water.
5.4
Transport
Activate the cleaning mode. Thus, the instrument will not interpret any
contact with the touch screen as an entry while being transported, e. g. in a
pocket (see chapter 5.3).
5.5
Disposal
Dispose of the meter without the battery and the sensor module as electronic
waste at an appropriate collection point.
Contamination
Cleaning agent
Reaction time at
room tempera-
ture
Water-soluble substances
Deionized water
Any
Fats and oils
Warm water and
household washing-up
liquid
Any
Heavy contamination
Methylated spirits
Max. 5 minutes
Lime and hydroxide coatings
Acetic acid (10 %)
Any