BYK LC 4
What to do if...
47
6
What to do if...
Error message
0FL
0FL
0FL
0FL, UFL
UFL
UFL
UFL
Error message
Err3
Err3
Err3
Err3
Error message
Err4
Err4
Err4
Err4
Error message
Err5
Err5
Err5
Err5
Cause
Remedy
– The measured value lies outside
the measuring range
– Sensor module defective
– Replace sensor module
Cause
Remedy
– Erroneous calibration
– The data of the sensor module is
outside the specified characteris-
tics.
– Recalibrate
– Reset to default setting (see
– Sensor module contaminated
– Clean and replace sensor
module if necessary
– Moisture in the plug connection
– Dry plug and socket
– Calibration solution not suitable
– Check calibration solution
Cause
Remedy
No sensor module
– Connect sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
– Replace sensor module
Cause
Remedy
Sensor module defective
or
Wrong sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
– Replace sensor module