1 processor reset, Processor reset – ALTANA LC 4 User Manual

Page 48

Advertising
background image

What to do if...

BYK LC 4

48

No stable

measured value

Indication,

l

Meter does not

react to keystroke

Sensor symbol

flashes

Message,

StoFull

StoFull

StoFull

StoFull

6.1

Processor reset:

The processor reset is a simple action that can be helpful in the case of error
messages or error modes.

Cause

Remedy

– Sensor module contaminated

– Clean sensor module

– Temperature not stable

– Adjust temperature if necessary

– Conductivity too low

– Use suitable sensor module

– Organic liquids

– Use suitable measuring system

Cause

Remedy

– Battery almost empty

– Exchange the battery (see

section 5.2)

Cause

Remedy

– Operating condition undefined or

EMC load unallowed

– Perform a processor reset (see

section 6.1)

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Clear the memory (see section

4.9.3)

1

Press <

1> and <8> at the same time and release them again.

Stored data and calibration values are retained.