1 processor reset, Processor reset – ALTANA LC 4 User Manual
Page 48
What to do if...
BYK LC 4
48
No stable
measured value
Indication,
l
Meter does not
react to keystroke
Sensor symbol
flashes
Message,
StoFull
StoFull
StoFull
StoFull
6.1
Processor reset:
The processor reset is a simple action that can be helpful in the case of error
messages or error modes.
Cause
Remedy
– Sensor module contaminated
– Clean sensor module
– Temperature not stable
– Adjust temperature if necessary
– Conductivity too low
– Use suitable sensor module
– Organic liquids
– Use suitable measuring system
Cause
Remedy
– Battery almost empty
– Exchange the battery (see
Cause
Remedy
– Operating condition undefined or
EMC load unallowed
– Perform a processor reset (see
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– All memory locations are full
– Clear the memory (see section
1
Press <
1> and <8> at the same time and release them again.
Stored data and calibration values are retained.