Buffalo Filter® VisiClear™ Operator’s Manual

Figure 4.

FOOTSWITCH / EZLINK AUTOMATIC ACTIVATION DEVICE

The VisiClear™ Surgical Smoke Plume Evacuator also comes equipped with a pneumatic

footswitch. A footswitch or an EZLink Automatic Activation Device may be added to any

system by simply plugging in a BUFFALO FILTER® activation accessory into the appropriate

jack on either the front (footswitch) or back (EZLink) panel of the unit. When the footswitch

is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for

each operation. For directions on using the EZLink Automatic Activation Device, please see

instructions that accompany that product.

FILTER LIFE INDICATOR

The filter life indicator on the touch screen control panel provides a visual indication of the

status of the life of the filter in use. The filter life indicator for the VisiClear™ Surgical Smoke

Plume Evacuator will automatically adjust according to the mode setting selected.

CAUTION: Using The filter life time and RFID Technology is a measure of time only, not a

measure of filter performance.

Laparoscopic Mode = 35 hours of filter life

Electrosurgical/Diathermy Mode = 21 hours of filter life

Open Tubing Mode = 18 hours of filter life

The VisiClear™ Filter Life Indicator is factory set. All filter life timing is automatic. See Figure 5.