Buffalo Filter VisiClear User Manual

Page 25

Advertising
background image

25

Buffalo Filter® VisiClear™ Operator’s Manual

Table 4

Recommended Separation Distance Between

Portable and Mobile RF Communications Equipment and the Model @ 3 Vrms

The model VisiClear™ is intended for use in the electromagnetic environment in which radiated RF

disturbances are controlled. The customer or the user of the model VisiClear™ can help prevent electromagnetic

interference by maintaining a minimum distance between portable and mobile RF communications equipment

(transmitters) and the model VisiClear™ as recommended below, according to the maximum output power of the

communications equipment.

Rated maximum

output power of

transmitter W

Separation distance according to frequency of transmitter m

150 kHz to 80 MHz

3.5

d =

√P

v

1

80 kHz to 800 MHz

3.5

d =

√P

v

1

800 kHz to 2.5 GHz

3.5

d =

√P

v

1

0.01

0.12

0.12

0.23

0.1

0.34

0.34

0.74

1

1.7

1.7

2.3

10

3.7

3.7

7.4

100

11.7

11.7

23.3

For transmitters rated at a maximum output power not listed above, the recommended separation distance (D)

in meters (M) can be estimated using the equation applicable to the frequency of the transmitter, where (P) is the

maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structure, objects, and people.

Maintenance

Section 3.0

M

ai

nt

en

an

ce

Se

ct

io

n 3.0

[ ] [ ] [ ]

Advertising