Buffalo Filter® VisiClear™ Operator’s Manual

Figure 5.

Filter life can be displayed in both hours and minutes or as a percentage using the Settings

button located in the lower left section of the touch screen. See Figures 6 and 7.

Figure 6.

Figure 7.

When the maximum filter life is consumed, the following warning screen will appear indicating

that a filter replacement is needed. After inserting a new VisiClear™ Filter, press the green

check mark to clear the warning and resume normal operation. During operation, if a filter

expires the system will not turn off but will remain running until the system is powered down

or 6 hours passes, whichever comes first. At that point, the VisiClear™ will no longer operate

until a new filter is inserted into the system. It is always best practice to replace an expired

filter immediately; however, a validated contingency life is built into the filter to ensure no

interruption to the operative procedure. See Figure 8.