Buffalo Filter VisiClear User Manual

Page 15

Advertising
background image

15

Buffalo Filter® VisiClear™ Operator’s Manual

The unit will continue to cycle the suction, checking to see if the obstruction has been cleared.

If the obstruction is cleared, the alarm will automatically turn off and the system will resume

normal operation. The user also has the option of disabling the alarm during continued use.

If the obstruction is not removed or you wish to disable this feature, you may disable the alarm

by pressing the following button on the alarm screen until the red X appears on the occlusion

symbol as shown. See Figure 11.

Figure 11.

ESU PENCIL MODE:

When the system is in an ESU Pencil mode and suction setting is 50% or higher, the unit will

alarm and the suction will stop if obstructed by an object. If an obstruction occurs, the system

will alarm and the following screen will display. See Figure 12.

Figure 12.

You have the option of clearing the obstruction (recommended) or disabling the alarm. Please

note: If the obstruction is cleared in ESU Pencil mode, the alarm will not automatically turn

off. Once the obstruction is removed, to reactivate the unit to normal operation, please press

following button on the alarm screen. See Figure 13.

Figure 13.