Buffalo Filter® VisiClear™ Operator’s Manual

3.4 Troubleshooting the System – see below.

PROBLEM

POTENTIAL CAUSE

CORRECTIVE ACTION

1. Smoke Evacuation

System is ON but

suction is minimal

or none.

1. Filter is not seated

completely.

2. Filter is clogged.

3. Vacuum hose or tube is

clogged.

4. Motor/blower is

obstructed.

1. Re-install VisiClear™ Filter, press firmly into place

and fully engage clip.

2. Replace filter with a genuine Buffalo Filter

VisiClear™ Filter.

3. Replace vacuum hose or tube with genuine

Buffalo Filter products.

4. Call BioMed or BUFFALO FILTER® Technical

Services at 1.800.343.2324 or 716.835.7000.

2. Smoke Evacuation

System does not

function even though

suction ON button is

depressed.

1. Not plugged into an

electrical outlet.

2. Fuses are blown.

3. Electronic system

failure.

4. Filter life has expired or

invalid filter inserted.

1. Check power outlet and connection to rear or side

panel of the machine.

2./3. Call BioMed or BUFFALO FILTER® Technical

Services at 1.800.343.2324 or 716.835.7000.

4. Replace filter with a genuine Buffalo Filter

VisiClear™ Filter.