Buffalo Filter® VisiClear™ Operator’s Manual

Table 2

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The model VisiClear™ is intended for use in the electromagnetic environment specified below.

The customer or user of the model VisiClear™ should assure that it is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level Electromagnetic Environment - Guidance

Electromagnetic

discharge (ESD)

IEC 61000-4-2

+6 kV contact

+8 kV air

+6 kV contact

+8 kV air

Floors should be wood, concrete or ceramic

tile. If floors are covered with synthetic

material, the relative humidity should be at

least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

+2 kV for power

supply lines

+1 kV for input/

output lines

+2 kV for power

supply lines

+1 kV for input/

output lines

Mains power quality should be that

of a typical commercial or hospital

environment.

Surge

IEC 61000-4-5

+1 kV differential

mode

+2 kV common

mode

+1 kV differential

mode

+2 kV common

mode

Mains power quality should be that

of a typical commercial or hospital

environment.

Voltage dips, short

interruptions, and

voltage variations on

power supply input

lines.

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

Mains power quality should be that

of a typical commercial or hospital

environment. If the user of the Model

VisiClear™ requires continued operation

during power mains interruptions, it is

recommended that the Model VisiClear™

be powered from an uniterruptible power

supply or a battery.

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Power frequency

(50/60 Hz) magnetic

field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should

be at levels characteristic of a typical

location in a typical commercial or hospital

environment.