Install the density kit

 Disconnect scale from power supply.
 Remove the standard weighing plate and replace it with the density kit.

!

Model EMB 2000-2V

Ensuring correct positioning, see illustration below.

 Place the platform for glass containers in a way that it does not touch the

weighing plate.

 Place beaker in the centre of the platform Make sure that it has no contact

with the frame.

 Hang the immersion basket on the rack. Make

sure that it is centred in the

recess.

 Pour the liquid into the glass beaker. Filling height should be approx. ¾ of the

capacity. Immerse thermometer

 Heat the liquid, the instruments or the displacement body until the

temperature is constant. Observe the warm-up time of the balance.

For more information, please refer to the operating manual supplied