KERN & SOHN EMB 500-1BE User Manual

Page 36

Advertising
background image

EMB-V-BA-e-1422

36

11 Uncertainty of Measurement for Density Determination of Solids


This table shows the approximate readability of the balance in connection with the
density kit. Observe that these values have only been determined by calculation and
that influent parameters such as described in chapter 6 have not been taken into
consideration.

Approximate display at density measurements

(when using a balance with a readability of 0.01g*)

Weight of sample

(g)





Density of sample
[g/cm

3

]

1

10

50

100

500

1000

2000

3000

4000

1

0.1 0.01 0.003 0.002 0.0005 0.0003 0.0003 0.0002 0.0002

3

0.4 0.04 0.01 0.005 0.001 0.001 0.0005 0.0004 0.0004

5

0.7 0.07 0.01 0.008 0.002 0.001 0.001 0.001 0.0006

8

1.2

0.1 0.02 0.01 0.003 0.002 0.001 0.001 0.001

10

1.5

0.1 0.03 0.02 0.004 0.002 0.001 0.001 0.001

12

1.7

0.2 0.04 0.02 0.004 0.002 0.002 0.001 0.001

20

2.9

0.3 0.06 0.03 0.01 0.004 0.003 0.002 0.002

*when using a balance with a readability of 0.1 g, the numbers in this table have to be
multiplied with 10. When using a balance with a readability of 0.001 g, divide the numbers
through 10.


Reading example for table:
In a balance with a resolution of 0.001 g and a sample with a weight of 10 g, whose
density is 5 g/cm

3

, the display graduation is at 0.007 g/cm

3

.

Advertising