8.3 Determining density of liquids

At the density determination of liquids, a sinker is used whose density is known. The
sinker is weighed first in air and then in the liquid whose density is to be determined.
From the weight difference results the buoyancy from where the software calculates
the density.

The density of the enclosed steel sinker can be determined as described in chapter,
8.2.1.
Or quickly and inexpensively in our calibration laboratory.
For further information please go to KERN- Hompage (

www.kern-sohn.com

).

8.3.1 Density determination while using the density kits KERN YDB-01

or YDB-02.

+

Install the density kit, see chapter 8.2.1

+

Mode to call up density determination of liquids.

1. Turn on the scale with

, "0.000" is displayed.

2. If the scale does not display, "0.000", press

3. Call up the density determination mode of liquids by pressing

.

"Liquid" is displayed briefly, followed by the actual entered density of the sinker.
Initial entry or if it is necessary to change them describes will be described later.

While using the same sinker, the entered density stays stored. Skip the following
steps for additional measurements and start by determining the density of the
liquid (step 7).