14 data output rs 232 c, Technical data, Pin allocation of balance output bushing – KERN & SOHN EMB 500-1BE User Manual

Page 44

Advertising
background image

EMB-V-BA-e-1422

44

14 Data output RS 232 C

Data output is carried out via interface RS 232 C.

The following conditions must be met to provide successful communication between
the weighing balance and the printer.

• Use a suitable cable to connect the weighing balance to the interface of the

printer/PC.
Faultless operation requires an adequate KERN interface cable (optional).

• Communication parameters (e.g. baud rate) of the scale and printer must

match.

1. Technical data

• 8-bit ASCII Code
• 1 start bit, 8 data bits, 1 stop bit, no parity bit
• Baud rate selectable at 1200, 2400, 4800, 9600 and 19200 Baud
• Miniature plug-in necessary (9 pole D-Sub)

2. Pin allocation of balance output bushing:

Front view:




Pin 2:

Transmit data

Pin 3:

Receive data

Pin 5:

Signal ground

Advertising