3. Explanation of the data transfer

rE Cr:

PRINT key

Press the PRINT key, at stable weight is transferred.

Remote control commands


The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.
Take into account that the following remote control commands must be sent without
a subsequent CR LF.

s

Function:

Stable weighing value for the weight is sent via the RS232
interface

w

Function:

Weighing value for the weight (stable or unstable) is sent
via the RS232 interface

t

Function:

No data are sent, the balance carries out the tare function.


a. Format for stable values
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B U

1

U

2

U

3

CR

LF


b. Format in case of fault
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

B B B B B B B B B B B E

r

r

o

r

CR

LF


c. Format for unstable values
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B B B B CR

LF


AU PC:
The weighing values are sent automatically and continuously, no matter if the value
is stable or unstable.


d. Format for stable values
1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B

U

1

U

2

U

3

CR

LF


e. Format in case of fault
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

B B B B B B B B B B B E

r

r

o

r

CR

LF


f. Format for unstable values
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B B B B CR

LF