KERN & SOHN EMB 500-1BE User Manual

Page 41

Advertising
background image

EMB-V-BA-e-1422

41

Selection
adjustment
weight




The adjustment weight can be selected out of four (CAL1 / Cal 2)
predetermined nominal values (ca. 1/4; 1/2; 3/4; Max). In order to
achieve high-quality weighing results in the sense of the
measuring technology, it is recommended to select the nominal
value as high as possible

 Keep

pressed in the weighing mode until [Pr] is

displayed.

 Press

repeatedly until „CAL“ will be displayed.

 Acknowledge using

, the current setting is displayed.

 Select desired setting by

.

 Acknowledge selection by

.

Filter

With the filter settings, the scale can be tailored to the needs of
the specific applications or environmental conditions.

 Keep

pressed in the weighing mode until [Pr] is

displayed.

 Press

repeatedly until "FiltEr" is displayed.

 Acknowledge using

, the current setting is displayed.

 Select desired setting by

.

SIo*

slow / insensible

fast/sensitive

Hr.

FSt

 Acknowledge selection by

.