KERN & SOHN SFB 300K-1LM User Manual

Page 26

Advertising
background image

SFB-BA-e-1324

26

6.7.2

Non-verified models

 Call-up menu item P3 CALCalLiner, see chap. 6.5.1

 Confirm by

, the password query „Pn“ will be displayed.

 Press subsequently

,

,

or

,

,

.

Ensure that there are no objects on the weighing pan.

 Wait for stability display, then press

.

When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait

for stability display, then press

.

When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.

Wait for stability display, then press

.

When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait

for stability display, then press

.

 After a successful linearisation the balance will carry out a

self-test. Remove adjusting weight during selftest, balance
will return into weighing mode automatically.

Advertising
This manual is related to the following products: