KERN & SOHN SFB 300K-1LM User Manual

Page 50

Advertising
background image

SFB-BA-e-1324

50

10 Data output RS 232C (optional)

Weighing data can be issued according to menu settings either via the RS 232C

interface or by pressing

via the interface.


This data exchange is asynchronous using ASCII - Code.

The following conditions must be met to provide successful communication between
the weighing balance and the printer.

Use a suitable cable to connect the weighing balance to the interface of the

printer. Faultless operation requires an adequate KERN interface cable.

Communication parameters (baud rate, bits and parity) of balance and printer

must match. For detailed description of interface parameters see chap. 8,
menu block

P1 COM“ or „P2 COM“.

10.1 Technical Data

10.2 Technical data

Connection

9 pin d-subminiature bushing

Pin 2 input

Pin 3 output

Pin 5 signal earth

Baud rate

Optional 600/1200/2400/4800/9600

Parity

8 bits, no parity / 7 bits, even parity / 7 bits, odd parity

10.3 Printer mode

Printout examples (KERN YKB-01N):

Weighing

ST, GS

1.000kg

Advertising