6.8 Verification

General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially
verified if they are used as follows (legally controlled area):

a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the

medical and pharmaceutical laboratory.

c) For official purpose.
d) For manufacturing final packages.


In cases of doubt, please contact your local trade in standard.

Verification notes:

An EU Qualification Approval is in existence for verified weighing systems. If a
balance is used where obligation to verify exists as described above, it must be
verified and re-verified at regular intervals.
Reverification is carried out according to the relevant national statutory regulations.
The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!

Verification of the weighing system is invalid without the "seal".


Notes on verified models

Access to conductor plate:

Remove seal
Open display unit

The application of the display unit as a weighing system able to be verified

requires that the contacts of the circuit board are short-circuited with the help
of a jumper [K1]. For non verifiable models remove the jumper.

To adjust, short-circuit the contacts of the circuit board, using a jumper [K2].

[K1]

[K2]